| Digital Caliper | Outside, inside, depth, and step dimensions | 0–150 mm or 0–300 mm | About ±0.02–0.05 mm | 0.01 mm | Check jaw alignment, sliding movement, zero return, display, battery compartment, and corrosion. | 12 months or according to use and risk | Recent calibration certificate, zero check, and comparison with certified gauge blocks. |
| Micrometer | High-precision outside diameter or thickness | 0–25 mm per frame; larger ranges available | About ±0.002–0.01 mm | 0.001 mm or 0.0001 in | Inspect anvils, spindle, ratchet, lock, frame, thimble, and zero position. | 6–12 months for critical inspection work | Calibration results across several points, traceability statement, and repeatability test. |
| Dial or Digital Indicator | Small displacement, runout, alignment, and comparative measurements | 0–10 mm or 0–25 mm travel | About ±0.01–0.03 mm | 0.001–0.01 mm | Check spindle motion, stylus tip, bezel, mounting stem, backlash, and repeatability. | 12 months | Test report showing travel error, hysteresis, repeatability, and return-to-zero performance. |
| Steel Tape Measure | General linear distance | 0–5 m, 0–10 m, or 0–30 m | Usually ±1–2 mm per metre, depending on class | 1 mm | Inspect graduations, hook movement, blade deformation, locking mechanism, and retraction. | Verification every 12–24 months or after damage | Comparison against a certified length standard and visible, undamaged graduations. |
| Digital Multimeter | AC/DC voltage, current, resistance, and continuity | Up to 600–1,000 V; current ranges vary | DC voltage commonly ±0.5–1.0% of reading | 3½–6½ display digits | Check leads, probe insulation, display, selector switch, input protection, and fuse condition. | 12 months; shorter for safety-critical applications | Calibration certificate, electrical safety test, lead inspection, and verification at multiple ranges. |
| Pressure Gauge | Gauge or absolute pressure | 0–1 bar to 0–1,000 bar, depending on design | About ±0.5–2.0% of full scale | Typically 0.1–1% of full scale | Inspect case, lens, dial, threads, pointer movement, leaks, and evidence of overpressure. | 12 months; more frequently in pulsating systems | Pressure-point calibration report, leak test, zero check, and correct connection size. |
| Infrared Thermometer | Non-contact surface temperature | Approximately −50 to 550 °C for common models | About ±1–2% of reading or ±2 °C | 0.1 °C | Inspect lens, trigger, display, battery contacts, emissivity settings, and optical damage. | 12 months or after exposure to extreme conditions | Verification against a suitable temperature reference with documented emissivity and distance settings. |
| Digital Thermometer with Probe | Contact temperature | Approximately −50 to 250 °C | About ±0.3–1.0 °C | 0.1 °C | Check probe cable, tip, insulation, connector, display, and response time. | 12 months | Calibration points covering the intended operating range and confirmation of probe identification. |
| Humidity and Temperature Meter | Relative humidity and ambient temperature | 0–100% RH; approximately −20 to 60 °C | About ±2–5% RH; ±0.5 °C temperature | 0.1% RH; 0.1 °C | Inspect sensor grille, response, display, battery contacts, and signs of condensation. | 6–12 months | Humidity and temperature test points, sensor condition report, and environmental stabilization record. |
| Illuminance Meter | Light level | Approximately 0–200,000 lux | About ±3–8% of reading | 1 lux or 0.1 lux on low ranges | Inspect photodiode cover, cable, display, range switch, and response consistency. | 12 months | Comparison with a calibrated light source and test results at low, medium, and high ranges. |
| Sound Level Meter | Sound pressure level | Approximately 30–130 dB | Typically ±1–2 dB | 0.1 dB | Inspect microphone capsule, windscreen, display, weighting controls, and response settings. | 12 months | Acoustic calibrator check before and after use, plus a laboratory calibration certificate. |